Product pricing advice
export@macylab.com
Emergency contact
86-10-15000509891
Leave a message for us
Online Message
Mobile phone browsing
Model:AA-1800H
Dowload:Download the attachment immediately
Consultation: Consultation
Contact: Contact
Solutions: Solutions
Mainframe
|
||
Model
|
:
|
AA-1800H
|
Light Source
|
:
|
≤6 lamps automatic turret, automatic alignment
|
Power Supply
|
:
|
110/220V (+5% ~ -10%), 60/50Hz; 5000VA
|
Lamp Current
|
:
|
pulsed power supply
|
Optical System
|
:
|
large 1800 /mm grating ruling, full closed optical system
|
Wavelength Range
|
:
|
190nm-900nm
|
Automatically peak find,a key opticaloptimization function
|
||
Wavelength Accuracy
|
:
|
≤0.15nm
|
Wavelength Repeatability
|
:
|
±0.1nm
|
Spectral Bandwidth
|
:
|
0.1nm、0.2nm、0.4nm、1.0nm、2.0nm
|
(5 steps with automatic changeover)
|
||
Baseline Stability
|
:
|
≤±0.002A/30 min (Static)
|
≤±0.005A/30min (Dynamic)
|
||
Absorbance Range
|
:
|
0-4A
|
Flame Analytical System
|
||
Detector
|
:
|
imported photomultiplier tube
|
Burner Head
|
:
|
full titanium combustion head, 50mm or 100mm generalcombustion head
|
Atomization Chamber
|
:
|
polymer explosion-proof spray chamber
|
Nebulizer
|
:
|
atomizer efficient glass atomizer, can also be customized
|
Ignition Type
|
:
|
microcomputer control, automatic ignition
|
Gas Control
|
:
|
automatic gas control system
|
Detection Limits(Cu)
|
:
|
0.002μg/mL
|
Precision
|
:
|
RSD≤0.5%
|
Graphite Furnace Analytical System
|
||
Heating Mode
|
:
|
vertical heating
|
Temperature Control Method
|
:
|
vertical optical temperature monitoring graphite tube wall temperature
|
Temperature Range
|
:
|
room temperature to 3000℃
|
The Program
|
:
|
automatic temperature control up to 20 order
|
Temperature Control
|
:
|
the furnace enriched up to 20 times
|
Characteristics Volume
|
:
|
0.5 × 10-12g (Cd)
|
Detection Limit
|
:
|
0.4 × 10-12g (Cd)
|
Precision
|
:
|
RSD≤3%
|
The Cooling Water
|
:
|
can choose cooling water circulation system
|
Safety
|
:
|
the graphite tube damage, water flow
|
air pressure and other alarmtemperature overheating protection
|
||
Autosampler Graphite Furnace
|
||
Sample Tray
|
:
|
70 sample cups, 6 reagent cup
|
Sample Volume
|
:
|
1-100μl
|
The Smallest Lincrement
|
:
|
0.1 μl
|
The Volume of Sample
|
:
|
better than 1%(at the time of 10mL-100mL)
|
Repeatability Volume
|
||
Repeated Sampling Frequency
|
:
|
up to 99 Times
|
Cleaning and Waste
|
:
|
each 500mL
|
Container Volume
|