ICP-6810 in the nonferrous gold industry
ICP6810 in the nonferrous gold industry
●The ICP6810 profile is introduced
The ICP6810 full-spectrum direct-read inductance-coupled plasma emission spectrometer, with excellent resolution, test accuracy and precision, is widely used in the rare earth industry, chemical industry, andOre analysis, metal smelting, geological research, drug safety, experimental research, environmental testing, food safety and other industries.Especially for the determination of elements in the ICP6810 zinc oxide industry.
picture1.ICP6810
Table 1. Main technical parameters of ICP6810
●ICP6810 main technical parameters
High frequency ionizer |
|
Working frequency |
27.12MHz |
Frequency stability |
﹤0.05% |
Output power |
800W ~1600W |
Output power stability |
≤0.05% |
OMatching way |
Automatic matching |
Scan the light divider |
|
light path |
Full reflection imaging light path, quartz prism two-dimensional dispersion system |
focal distance |
400mm |
Graster specifications |
Middle step grating, 52.67, l p / mm, 64 shining angle |
Resolution ratio |
<0.0068nm@200nm |
Complete machine technical indicators |
|
Scan the wavelength range |
At 165-900nm, full wavelength coverage, and measurable Al 167.120nm |
Repetitiveness |
RSD≤1.5% |
Stability |
RSD≤2.0% |
●ICP6810 element detection limit (g / L)
element |
La |
Ce |
Pr |
Nd |
Sm |
Eu |
Gd |
Tb |
wavelength (nm) |
408.672 |
413.765 |
414.311 |
401.225 |
360.946 |
381.967 |
342.247 |
350.917 |
detection limit |
<3.0 |
<5.0 |
<5.0 |
<5.0 |
<10.0 |
<1.0 |
<10.0 |
<3.0 |
element |
Dy |
Ho |
Er |
Tm |
Yb |
Lu |
Y |
Sc |
wavelength (nm) |
353.170 |
345.600 |
337.271 |
313.126 |
369.419 |
261.541 |
371.030 |
335.373 |
detection limit |
<3.0 |
<3.0 |
<3.0 |
<3.0 |
<1.0 |
<3.0 |
<1.0 |
<1.0 |
element |
Ta |
Nb |
Mn |
Mg |
B |
Zn |
Co |
Si |
wavelength (nm) |
226.230 |
313.340 |
257.610 |
279.553 |
249.773 |
213.856 |
228.616 |
251.611 |
detection limit |
<5.0 |
<5.0 |
<3.0 |
<1.0 |
<10.0 |
<3.0 |
<3.0 |
<10.0 |
element |
Ni |
Cd |
Fe |
Ca |
Mo |
V |
Be |
Ti |
wavelength (nm) |
232.003 |
226.502 |
239.562 |
393.366 |
281.615 |
310.230 |
313.041 |
334.941 |
detection limit |
<5.0 |
<3.0 |
<3.0 |
<1.0 |
<5.0 |
<5.0 |
<1.0 |
<3.0 |
element |
Cu |
Cr |
Al |
Zr |
Ag |
Sr |
Au |
Pt |
wavelength (nm) |
324.754 |
267.716 |
396.152 |
343.823 |
328.068 |
407.771 |
242.795 |
265.945 |
detection limit |
<3.0 |
<5.0 |
<5.0 |
<5.0 |
<3.0 |
<1.0 |
<5.0 |
<5.0 |
element |
Pd |
Ir |
Rh |
Ru |
Ba |
Li |
Na |
K |
wavelength (nm) |
340.458 |
224.268 |
343.489 |
240.272 |
455.403 |
670.784 |
588.995 |
766.490 |
detection limit |
<5.0 |
<10.0 |
<10.0 |
<5.0 |
<1.0 |
≤3 |
≤20 |
≤60 |
element |
As |
Sb |
Bi |
Hg |
Pb |
Ga |
Os |
W |
wavelength (nm) |
228.812 |
206.833 |
223.061 |
253.652 |
220.353 |
294.364 |
225.585 |
207.911 |
detection limit |
≤15 |
≤15 |
≤10 |
≤15 |
≤15 |
≤10 |
≤1 |
≤10 |
element |
Sn |
Te |
Ta |
Th |
Tl |
Re |
Ge |
Se |
wavelength (nm) |
242.949 |
214.281 |
226.230 |
283.730 |
276.787 |
227.525 |
209.426 |
203.985 |
detection limit |
≤20 |
≤10 |
≤5.0 |
≤10 |
≤30 |
≤5 |
≤15 |
≤30 |
●Evaluation of alloy measurable elements in non-ferrous precious metals
For precious metal samples and customer test requirements, the details are shown in the Table:
2. Precious metal sample test solution
sample |
measureelement |
Analysis instrument |
Sample handling overview (for reference only) |
remarks |
Non-ferrous precious metal alloy |
Gold, silver, palladium, platinum, and so on |
ICP6810 |
Take 0.1-0.3g sample in a 100ml beaker, add 5ml hydrochloric acid, dissolve the sample on the heating plate to clarification, remove cold to room temperature, transferred to 50ml volumetric flask diluted with water to scale for testing. |
Sample blank experiment was also performed simultaneously duringthesampleprocessing process. |
According to the configured standard solution concentration, directly on the computer test, the instrument can automatically give the final test results, and can be directly printed in the form of a report.
Table 3.ICP6810 and related instrument configuration
Instrument host and necessary accessories |
||||||
order number |
definition |
Specifications / models |
manufacturer |
quantity remarks |
quantity remarks |
|
1 |
Full-spectrum straight-read inductively coupled plasma emission spectrometer |
ICP6810 |
Shanghai Beauty Instrument Co., Ltd |
1 |
|
|
2 |
Automatic temperature control and circulating cooling water device |
|
|
1 |
|
|
3 |
stabilized voltage supply |
|
|
1 |
|
|
Instrument matching accessories |
||||||
order number |
definition |
Specifications / models |
manufacturer |
quantity remarks |
quantity remarks |
|
1 |
host |
|
|
1 |
|
|
2 |
graphoscope |
|
|
1 |
|
|
3 |
marking punch |
|
|
1 |
|
|
Sample processing equipment and consumables |
||||||
order number |
definition |
Specifications / models |
manufacturer |
quantity remarks |
quantity remarks |
|
1 |
Electronictemperature-controlled heating board |
EG-35B |
|
1 |
|
|
2 |
One in ten thousand is an electronic balance |
BS224S |
|
1 |
|
|
3 |
Laboratory-grade ultra-pure water machine |
EPED-40T |
|
1 |
|
|
4 |
Disposable rubber Gloves |
M |
|
/ |
|
|
5 |
Activated carbon mask |
|
|
/ |
|
|
6 |
3mlpasteurpipette |
3ml |
|
/ |
|
|
7 |
100mlglass beaker |
100ml |
Sichuan Shu Glass scientific Instruments |
/ |
|
|
8 |
100mlGlass capacity bottle |
100ml |
Sichuan Shu Glass scientific Instruments |
/ |
|
|
9 |
5mltransfer liquid gun |
0.5ml-5ml |
Rand, Germany |
1 |
|
|
10 |
1mltransfer liquid gun |
100ul-1000ul |
Rand, Germany |
1 |
|
|
Reagent |
||||||
order number |
definition |
Specifications / models |
manufacturer |
quantity remarks |
quantity remarks |
|
1 |
hydrochloric acid GR Jiangsu Qiangsheng Chemical Co., Ltd |
hydrochloric acid GR Jiangsu Qiangsheng Chemical Co., Ltd |
hydrochloric acid GR Jiangsu Qiangsheng Chemical Co., Ltd |
/ |
|
|
2 |
hydrogen nitrate GR Jiangsu Qiangsheng Chemical Co., Ltd |
hydrogen nitrate GR Jiangsu Qiangsheng Chemical Co., Ltd |
hydrogen nitrate GR Jiangsu Qiangsheng Chemical Co., Ltd |
/ |
|
|
3 |
perhydrol GR Jiangsu Qiangsheng Chemical Co., Ltd |
perhydrol GR Jiangsu Qiangsheng Chemical Co., Ltd |
perhydrol GR Jiangsu Qiangsheng Chemical Co., Ltd |
/ |
|
|
4 |
Standard material of each element 1000ug/ml National nonferrous Metals and electronic materials center of forecasting and analysis |
Standard material of each element 1000ug/ml National nonferrous Metals and electronic materials center of forecasting and analysis |
Standard material of each element 1000ug/ml National nonferrous Metals and electronic materials center of forecasting and analysis |
/ |
|
|
If you have any other questions, please call the analysis instrument